X-Ray High Voltage Cables
Configuration
Cable Length | ||
Standard | 8m | 12m |
Customized | 15m | 20m |
Specification
Electrical Data | |
Rated Voltage | 90kVdc |
Shield Resistance | 0.010 Ohm/m |
Coverage Shielding Braid | > 95% |
Cable Capacity | < 120pF/m |
Test Voltages | |
High Voltage Insulation | 90kVdc |
Low Voltage Insulation | 500 Vac |
Mechanical Data: Cable | |
Outer Diameter | 16.5 mm + 0.5 |
Bending Radius | > 150mm |
Low Voltage Conductor | 2×1.5mm2 |
Common Conductor | 2×0.75mm2 |
Conductor Resistance | 0.013 Ohm/m |
Cable Sheat Colour | Light grey |
Mechanical Data: Plug | |
Multiple Contact Pin | 8 point contact |
Max. Temperature | 110 ℃ |
Flanges | Demountable |
Extremely Short Sleeve | 34 mm |
Small Plug Diameter | 40 mm |
Solderless Connection | (EMC grounding) |
The high voltage cables are mainly used on large and midium scale x-ray
machine, which is connected with the high voltage generator and x-ray tube
to transfer the high voltage from generator to the anode and cathode of the
x-ray tube, and transfer the heating voltage to the x-ray tube filament.